A Rare Case of Acute Stent Recoil in Acute Inferior Wall Myocardial Infarction Patients

Dr. Basavaraj Bramhanand Sutar1*, Dr. Prabhakaran Aranganathan2, Mr. Manivasagam3, Dr. Paulraj I S Chakravarthi4
 

1. Dr. Basavaraj Bramhanand Sutar, MBBS; DNB (Med); DNB cardiology.

2. Dr. Prabhakaran Aranganathan, MBBS; DNB (Med); DNB cardiology.

3. Mr. Manivasagam. S, Diploma CVT.

4. Dr. Paulraj I S Chakravarthi, MD, MRCP(UK), CSCST(Cardiology).

*Corresponding Author: Dr. Basavaraj B. Sutar. Department of Cardiology, GKNM hospital, Pappanaickenpalayam, Coimbatore, Tamil nadu, India- 641037.


Received Date:  March 18, 2021

Publication Date: April 01, 2021

DOI:10.1027/marcy.2021.0138

 

Abstract

Acute stent recoil is a rare phenomenon that can precipitate acute stent thrombosis or in-stent restenosis. Factors associated are reduced strut thickness, a larger stent/vessel ratio, and a larger balloon/stent ratio. We report a case of acute stent recoil of Promus Elite, a platinum chromium everolimus-eluting stent in a 61-year-old male who underwent Rescue PCI to RCA. Acute stent recoil was noted after post dilatation which was tackled by deploying another Resolute Onyx, a platinum-iridium zotarolimus-eluting stent. Acute radial collapse probably occurred due to inadequate lesion preparation, high-pressure post dilatation with a larger balloon, and relatively thin struts of the stent.

Keywords: Acute stent recoil, rescue PCI, ST-elevation Myocardial infarction, DES.


A Rare Case of Acute Stent Recoil in Acute Inferior Wall Myocardial Infarction Patients

Text of manuscript

Case report

History of presentation/ Past Medical History/Investigations

A 61-year-old male with a background history of Diabetes and Hypertension presented to the Emergency room with chest pain of 4 hours duration. He was diagnosed to have Acute coronary syndrome – Inferior wall myocardial infarction. ECG showed ST elevation in lead II III and save. ECHO revealed RWMA and EF 51%. He was hemodynamically stable and was thrombolysis with Tenecteplase. The patient had Ventricular tachycardia post thrombolysis from which he was resuscitated and intubated. Given persistent ST elevation, ventricular arrhythmias, he was taken up for rescue PCI. TPI was done through the Right femoral approach. CAG done through right radial approach revealed Two vessel coronary artery disease and proximally occluded RCA (Fig.1) was the cause of recent Inferior wall Myocardial infarction.

Management

RCA was engaged with 6F Judkins Right guiding catheter and wired with 0.014” Sion blue guide wire when the flow was established. Proximal RCA had 75% thrombus containing lesion which was directly stented with 3 X 38 mm Promus Elite platinum chromium everolimus-eluting stent at 10 atm. (Fig 2) . Post dilatation of the stent was done with 3.5 X 15 mm NC Quantum apex noncompliant balloon at 16 atm distally and at 20 atm proximally.

Even though the stent expanded well during deployment and post dilatation, acute stent recoil was noted after few minutes (Fig 3).

Post dilatation with 3.5 x 15 mm non-compliant balloon was done repeatedly at the recoiled segment upto18 atm. However, the segment repeatedly recoiled even after many dilatations precipitating slow flow and hemodynamic instability ( Fig 5, 6 & 7). Another 3.5 x 22 mm Resolute Onyx platinum-iridium zotarolimus-eluting stent was deployed covering the recoiled segment at 12 atm for 10 sec (Fig 7,8 & 9). Post dilatation was done with a shorter 3.5 x 8 mm NC Quantum apex non-compliant balloon at 20 atm.

The recoiled segment expanded well and no angiographical residual stenosis was observed. IVUS was not done given hemodynamic instability due to recoiling. TIMI III flow was achieved with no procedural complications.

 

Discussion

Acute stent recoil is a rare phenomenon with an estimated incidence of 1.2% among first or second-generation stents (1) that may lead to acute stent thrombosis or stent restenosis which have serious adverse outcomes. There are multiple factors postulated such as reduced strut thickness, a large stent vessel ratio, a large balloon stent ratio, aggressive high-pressure balloon dilatations. Platinum chromium alloy has been used as a backbone in coronary stents recently for its reduced thickness and better deliverability. Koo et al (2) reported that reduced strut thickness is associated with decreased radial strength and more acute stent recoil. However platinum chromium stents are found to be non-inferior as compared with similar cobalt-chromium alloy stents (3). Stent strut thickness alone may not be the reason for recoil in our patient.

We considered lack of lesion preparation could be a possible reason but the stent expanded well during deployment and initial post dilatation. Recoil was noted after few minutes suggesting that inadequate lesion preparation might not be contributory.

Bommel et al (4) reported large stent vessel ratio and large balloon stent ratio are predictors of acute stent recoil. We used a larger balloon 3.5 mm for a 3 mm stent in our patient resulting in a balloon stent ratio >1. Takagi et al (5) suggested that high-pressure dilatations may lead to acute stent recoil. Even though we inflated up to 16 atm initially at the distal segment, we inflated up to 18 atm after recoil aiming at good expansion.

 We ended up deploying another stent in recoiled segment followed by post dilatation thereby having increased strut thickness to achieve good expansion. In our patient, stent strut thickness and inadequate lesion preparation might have not been contributory. We conclude high-pressure dilatation with an oversized balloon as the reason behind acute stent recoil in our patient. However, the sizing was done based on an angiogram which may not be accurate. We did not do IVUS given hemodynamic instability. The ideal would be doing imaging for all possible cases to choose appropriate stent and balloon and using prolonged inflation during deployment, multiple optimal pressure dilatations for good stent expansion rather than using an oversized balloon and high-pressure dilatations.

 

Follow-up

During subsequent LAD stenting after 72 hours, check CAG revealed RCA stents with no residual stenosis.

 

Conclusion

Even though thin struts are expected to recoil more, platinum chromium alloy does have good radial strength and stent recoil is multifactorial. Imaging guided appropriate stent and balloon selection, multiple optimal pressure dilatations for adequate stent expansion would help prevent acute stent recoil. High-pressure dilatations of an oversized balloon should be avoided. A double stenting strategycan be considered as a bailout in acute stent recoil.

 

Learning objectives

    1. Multiple optimal pressure dilatations for adequate stent expansion would be helpful in preventing acute stent recoil.

    2. High pressure dilations of oversized balloon should be avoided and double stenting strategy can be helpful in acute stent recoil, imaging can be more helpful to assess stent recoil which is limitation of our case report.

 

References

1. Ohya M, Kadota K, Kubo S, et al. “Incidence, predictive factors, and clinical impact of stent recoil in stent fracture lesion after drug-eluting stent implantation”. Int J Cardiol. 2016; 214: 123– 129, doi: 10.1016/j. ijcard.2016.03.013, indexed in Pubmed: 27060271.

 

2. Koo BK, Waseda K, Ako J, et al. “Incidence of diffuse and focal chronic stent recoil after implantation of current generation bare- -metal and drug-eluting stents”. Int J Cardiol. 2010; 144(1): 132–134, doi: 10.1016/j.ijcard.2008.12.117, indexed in Pubmed: 19171393.

 

3. Park, K. W., Kang, S.-H., Kang, H.-J, et al. (2014). A Randomized Comparison of Platinum Chromium-Based Everolimus-Eluting Stents Versus Cobalt Chromium-Based Zotarolimus- Eluting Stents in All-Comers Receiving Percutaneous Coronary Intervention. Journal of the American College of Cardiology, 63(25), 2805–2816. https://doi.org/10.1016/j.jacc.2014.04.013.

 

4. Bommel Rv, Lemmert M, Mieghem Nv, et al. “Occurrence and predictors of acute stent recoil- A comparison between the xience prime cobalt chromium stent and the promus premier platinum chromium stent”. Catheter Cardiovasc Interv. 2017; 91(3): E21–E28, doi: 10.1002/ccd.27096.

 

5. Takagi K, Yakushiij T, Basavarajaiah S, et al. “Acute stent recoil from aggressive post-dilation of a second-generation drug-eluting stent”. JACC Cardiovasc Interv. 2013; 6(3): 311–313, doi: 10.1016/j. jcin.2012.08.027, indexed in Pubmed: 23517845.

 

Volume 2 Issue 4 April 2021

©All rights reserved by Dr. Basavaraj B. Sutar.

Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

Figure 8

rm pola mahjong abc mw2rm pertunjukkan istimewa sifupola menang mw3rahasia dapurrahasia wildramuan ajaib mahjongrasakan sensasi bedabonus new memberbosan miskin tiapbuktikan kamu pantascara cerdas bermaincoba pola inimakin bosan kehidupanmahjong ways jarang10 trik rahasiaanalisa pola mahjongapa yang membuatawas ketinggalan rahasiabagi yang inginkeseruan nonstopkunci kemenaanganmahjong ways 2rm paman herry jackpot dadakanrm perjalanan bg fred0878 tahun baru 20250879 tambah gacor0880 mahjong teknik mantul0881 olympus zeus marah0882 pola gacor full senyum0883 mahjong 7 tips0884 strategi game gachor0885 game pilihan terbaikrtp gacor olympusscatter bonanza jackpot hingga ratusan jutagame online pragmatic pecah jackpotjam tepat buy spin mahjong waysputaran maxwin starlight princesskejutan besar game mahjongpola ghacor mahjong winsmahjong wins freebetgame mahjong online terbesarbanjir scatter game mahjong waysserver game luar negeritrik pragmatic play gampang menanggame olympus pintu kekayaanrm naga baru naga berlainrm om kenny hidup mewahrm paling gampang jackpot brocara betting hemat mahjong ways 2cara menang mahjong wins 3kombinasi pola mahjong ways 2optimalkan kemenangan dengan rtp livepanduan menang sweet bonanzakomunitas mahjong ways 2cara menang starlight princess 1000keuntungan mahjong ways server thailandpola kemenangan mahjong ways 2pola terbaru lucky neko 2025badai petir kakek zeus di gates of olympuscara memunculkan scatter hitamdaftar akun vip auto maxwinnaga hitam mahjong pg softrahasia simbol scatter dan wild mahjong ways 2akun vip mahjong ways resmibocoran pg soft anti boncosgame ghacor pragmatic play terbarupola full scatter gates of olympuspola sakti starlight princessrahasia rtp tertinggi gates of olympusscatter naga hitam mahjong winstips maxwin game Olympus untuk pemulatrik maxwin mahjong ways terbarutrik sweet bonanza anti rungkadrahasia sukses mahjong ways 2spesial tahun baru mahjong ways 2 scatter hitamstrategi cuan starlight princess x1000strategi menang mahjong waystahun baru 2025 mahjong wayspromosi kincir88 gates of olympus mahjong waysscatter emasscatter hitam mahjong waysscatter hitam mimpi wede besarstrategi maxwin mahjong ways 2025rtp live mahjong ways pg softsweet bonanza x1000 tanpa polatips spin otomatis scatter hitam mahjong waystrik scatter hitam akun vvip mahjong wayswinrate tinggi scatter hitamjackpot gilapaman herryrahasia para masterrtp tertinggi januarivolatilitas tinggi mahjongrm kuasai alam semsta mahjongrm melukis jackpot tiada yg mustahilrm menang situs rm1131 gampanghoki menggilailmu sihir mahjongjackpot melimpahkenali multiplierpundi pundimultiplier gilanenek ani jackpotbocoran pola mahjong winsmahjong waysmaxwin sweet bonanzartp maxwin terbarucara membaca rtp mahjong ways 2gates of olympus rtp live 2025maxwin scatter hitam mahjong wayspola kemenangan mahjong wins3pola rahasia mahjong waysgates of olympus modal minimmahjong hitam di adat88rtp mahjong ways 2 terbarutrik jackpot di adat88budi sultandapati jackpotdominasi arena mahjonggacor abisgebyar kemenangankonsentrasi kuncikuasai kuncimaxwin bersama