Evaluation of the Structural, Flowmetric and Functional Effect of Panretinal Photocoagulation With 532-Nm Argon Laser Compared to 577-Nm Pattern Scanning Laser in Diabetic Retinopathy
Joaquín Sosa Lockward1, Nassim Abreu Arbaje2, Miguel Cruz Pimentel3, Juan F. Batlle4
1. 3rd-year ophthalmology resident, Dr. Elías Santana hospital, Santo Domingo, Dominican Republic
2. Associate Professor of the Retina Department, Dr. Elías Santana hospital, Santo Domingo, Dominican Republic
3. Senior Fellow of the Retina Department, Dr. Elías Santana hospital, Santo Domingo, Dominican Republic
4. Chairman of the Retina Dept, Dr. Elías Santana hospital, Santo Domingo, Dominican Republic
Corresponding Author: Dr. Joaquín Sosa Lockward, 3rd-year ophthalmology resident, Dr. Elías Santana hospital, Santo Domingo, Dominican Republic
CopyRight: © 2021 Joaquín Sosa Lockward, This is an open-access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Received Date: September 03, 2021
Published date: October 01, 2021
Abstract
Purpose: To compare the effect of 532-nm argon laser and 577-nm pattern scanning laser in the anatomy, vasculature and retinal function in patients with diabetic retinopathy
Methods: This was a single-center, interventional, single-blind case series study. Twelve patientes without macular edema and with clinical criteria for treatment according to ETDRS study underwent panretinal photocoagulation with 532-nm or 577-nm laser by a single retina specialist. Changes in retinal morphology and function were assessed before photocoagulation as well as the 1st week, 4th weeks, 8th weeks and 12th weeks after treatment.
Results: Twelve patients were enrolled in 532-nm laser treatment (50%) or 577-nm laser treatment (50%) homogeneously. No significant difference was seen in visual acuity (p = 0.39), contrast sensitivity (p = 0.76) and color vision (p = 0.21) 3 months after PRP. Additionally, the papillary vascular density was not different at the end of the follow up in both groups (p > 0.05) The average central retinal thickness (CRT) at the 8th week was higher in the argon group (287μm ± 59μm vs. 268μm± 24μm; p < 0.05). Furthermore, the initial size as well as the scar expansion was larger in the argon group in all visits (p < 0.001).
Conclusions: There were no significant changes in retinal function at 12th weeks after PRP. The papillary vascular density was restored at the end of the follow up. Patients who underwent argon laser therapy had a significant increase in CRT with larger scar expansion compared to pattern scanning laser group.
Keywords: Diabetic Retinopathy; Argon laser; Pattern scanning laser; Panretinal photocoagulation; Central retinal thickness; papillary vascular density.
Introduction
Diabetic retinopathy is the number one cause of visual impairment in working-age adults around the world. This disease is characterized by poor capillary perfusion and retinal ischemia, which leads to macular edema and neovascularization, and can severely alter the visual function of these patients.
Likewise, although panretinal photocoagulation (PRP) has provided an effective treatment to reduce the risk of severe visual loss in patients with diabetic retinopathy, it is not absent of complications that can result in central function loss due to macular edema, peripheral constriction of the visual field with poor adaptation to darkness and accommodation (1).
For that reason, different types of lasers have been created with modifications of international protocols to try to minimize these effects as much as possible and the PASCAL® laser has been postulated as a proposal to this objective, controlling energy delivery in short pulse duration (2).
The laser energy is transformed into heat when absorbed by the melanin of the retinal pigmented epithelium and the hemoglobin contained within the vessels, so with less pulse duration, less energy extends in all directions changing in a more innocuous way the retinal tissue, with less lateral thermal diffusion and damage to the RPE and less alteration of the internal retina, with a better patient tolerance and less collateral damage compared to the conventional laser treatment.(2)
Objectives
General objetive
To compare the effect of 532-nm argon laser and 577-nm pattern scanning laser in the anatomy, vasculature and retinal function in patients with diabetic retinopathy
Specifics objetives
Methods:
The current study was a single-center, interventional, single-blind case series study of patients of 18 year of age or older with severe, very severe or proliferative diabetic retinopathy based on the clinical criteria of the ETDRS.
Subjects were enrolled with informed consent between August 2019 and March 2020. Eyes with the following conditions were excluded: Severe cataract (Stated as NO3NC3C3SCP3 or worse following LOCS III classification), any grade of vitreous hemorrhage, CRT more than 300μm, history of PRP, any retinochoroid pathology other than diabetic retinopathy, history of intravitreal injection within the past 3 months and any optic neuropathy. Demographic data, ophthalmic examination findings and other relevant clinical data was extracted from patient personal record file.
All patients underwent a clinical examination that included baseline best corrected visual acuity (BCVA) with Snellen ETDRS letters, contrast sensitivity with Pelli-Robson chart (From 0 to 100%) and chromatic vision with standard ishihara test (14 color plates). Furthermore, a slit-lamp and dilated fundus examination was performed by a single ophthalmologist altogether with a baseline OCT (Optovue®) to quantify CRT and an OCT-A to measure the papillary vascular density using the platform angiovue (Optovue®).
After informed consent approval, enrolled patients underwent PRP using 532-nm single spot argon laser (OcuLight®,IRIDEX, Mountain View, CA) or 577-nm pattern scanning laser (PASCAL Synthesis® Topcon Canada Inc.) by another single retina specialist. Pupillary dilation was obtained with combined phenylephrine and tropicamide drop (T-P, 50mg/8mg/ml, Sophia®). Topical anesthesia with one drop of proparacaine (Anestears, Lansier®) was administered before laser treatment.
Every patient underwent two separate sessions spared with two weeks between them, using the following parameters: (1) Stardard 200μm spot size using a single lens (Volk® transequator), (2) power and duration were subject to variation until a gray light lesion was attained, thus documenting the average power used in both groups at the end of each laser session. (3) The spots were placed 1 spot away in the argon group and a 3x3 grid with 1.25μm spacing in the pattern scanning laser group. In the subjects, an inferotemporal area below the vascular arcade was selected to measure the scar expansion in no more than 3 scars.
Immediately after the treatment, total session time was recorded and patient perceived pain was assessed with a standardized subjective pain scale from 0 to 10, where 0 indicated that the subject experienced no pain and 10 indicated that the procedure was extremely painful. In that way, another single ophthalmologist performed the measurements of the other variables in the 1st week, 4th weeks, 8th weeks and 12th weeks after treatment.
All the data was analized on SPSS version 23. Significance level considered as p<0.05.
Results
Twelve eyes with severe, very severe or proliferative diabetic retinopathy based on the clinical criteria of the ETDRS were enrolled. The BCVA (logMAR) was 0.25 ± 0.10 before treatment, 0.28 ± 0.11 at the 4th week, 0.3 ± 0.11 at 8th weeks and 0.26 ± 0.11 at the 12th weeks in both de PASCAL and argon group. There was no significant difference between the groups at any time point. (p = 0.39)
(Figure 1.)
The contrast sensitivity was measured from 0 to 100%. Changes remained almost linear, and no statistical difference was noted from baseline nor between both laser treatment group at any time point using a friedman test (p = 0.76)
In terms of chromatic vision at 12th weeks in the argon group, 1 patient failed in one of the plates of the ishihara test, and another patient presented 2 failures or more. All the remaining patients asserted the 100% of the plates in all the visits. However, this data did not obtain statistical significance after performing a Kruskall-Wallis test for independent variables. (p = 0.21)
The subjective pain was measured from 0 to 10 immediately after panretinal photocoagulation treatment (Figure 2).
The mean pain scale grade was 8.5 ± 1.5 in the argon group and 6 ± 1.4 in the PASCAL group. These changes were statistically significant using a t-student test (p < 0.05).
Mean panretinal photocoagulation session time was 7.69 ± 5.30 minutes and 4.63 ± 2.63 minutes in the argon and PASCAL group respectively (p <0.05). The mean number of PRP burn was 657 ± 377 in the argon group and 887 ± 434 in the PASCAL group (p <0.05) (Figure 2).
The mean CRT was 256μm ± 23μm before laser treatment, 261μm ± 19μm at the 1st week, 262μm ± 20μm at the 4th week, 268μm ± 25μm at the 8th week and 265μm ± 25μm at the 12th week in the PASCAL group (Figure 3).
These changes were not statistically significant from baseline using a friendman test (p <0.05) In the argon group, the mean CRT was 259μm ± 28μm before laser treatment, 253μm ± 26μm at the 1st week, 268μm ± 32μm at the 4th week, 287μm ± 59μm at the 8th week and 274μm ± 43μm at the 12th weeks (Figure 4).
The authors found an important statistical difference at 8th weeks from baseline measure and between both argon and PASCAL groups (p <0.001). (Figure 5)
The mean peripapilar vascular density was 56% ± 1.27% before laser treatment, 55.24% ± 3.82% at the 1st week, 55.46% ± 3.48% at the 4th week, 56.28% ± 1.27% at the 8th week and 56.92% ± 2.08% in the 12th week in the argon group. (Figure 4) The PASCAL group presented 56% ± 1.45% before laser treatment, 54.56% ± 3.14% at the 1st week, 55.1% ± 3.51% at the 4th week, 53.18% ± 3.82% at the 8th week and 54.9% ± 4.11%. (Figure 3) There was not a statistical difference between baseline and the last follow up using a friedman test (p <0.05)
Regarding scar expansion, the mean scar horizontal diameter was 304μm ± 103μm in the 1st week after laser treatment, 288μm ± 54.6μm at the 4th week, 329μm ± 51μm at the 8th week and 354μm ± 46μm in the argon group. The PASCAL group presented a mean scar horizontal diameter of 162μm ± 46μm at the 1st week after treatment, 171μm ± 27μm at the 4th week, 198μm ± 32μm at the 8th weeks and 214μm ± 31μm at 12th weeks. There was an important statistical difference between the scar diameter in every visit between de argon and PASCAL laser treatment groups using a kruskall-Wallis test (p < 0.001) (Figure 6)
Discussion
The main objective of this study was to evaluate the changes in the anatomy, vasculature, and retinal function in patients with diabetic retinopathy with good visual acuity and no macular edema that were treated with PRP either with conventional argon laser or PASCAL laser. In our study, non statistically significant changes were noted in BCVA, chromatic vision or contrast sensitivity at the end of the follow up between both laser treatment group.
The subjective patient perceived pain was higher in the conventional argon laser group (8.5 ± 1.5 vs. 6 ± 1.4). Mirashi et al. compared pain scores in patients with severe NPDR or PDR, in which one eye of each patient was randomized to undergo conventional laser and the other one to short pulse laser. In their study, subjects had an average pain score of 7.5 ± 1.14 and 1.75 ± 0.87 in the conventional and in the short duration laser group respectively. (3) Reduced pulse duration may be correlated with less pain due to decreased thermal diffusion into the choroid, which is rich in sensory nerves, resulting in minor collateral damage. This fact associated with shorter laser sessions makes short pulse laser therapy more tolerable by patients as demonstrated in this study. (3-5)
In our data, patients in the conventional argon group presented an increase in CRT at the 8th week that had statistically significance. Many studies had evaluated this effect in CRT (3, 6-8) Mirashi et al. study demostrated an increase in mean central macular thickness as early as the 1st week (37.9 μm vs. 8.7μm) that remained 4 months later (50.8μm vs. 9.1μm) (p<0.05) (3) Takamura et al. showed that more patients in the conventional laser group had more than a 20% increase in CRT from baseline in comparison to the PASCAL group at 18th weeks (23.8% vs. 9.5%) (p<0.05) (7) Similar results were presented by Mahgoub et al. with a slight increase from baseline (33.68μm vs. 26.06μm) (8) This stated difference is probably associated with the fact that with a longer duration, there is a greater tissular disruption with subsequent release of IL-6, ICAM-1 and MCP-1 that upregulates VEGF and causes an increase in vascular permeability(1,7)
Other methods described previously have been used to quantify the variation in the optic nerve head flow after PRP, like speckle flowgraphy, and there is limited literature with the use of OCT angiography in this specific field. As far as the authors are aware, there’s no comparative data of the variation in the optic nerve head flow using this method between conventional argon laser and PASCAL laser photocoagulation. In our study, there was a slight diminish in the optic nerve head flow in the 4th week and the 8th week in both groups respectively (Figure 7).
At the end of the follow up, the ONH flow was not different from the baseline measure, suggesting recovery as showed by Iwase et al. (9) The possible explanation is that PRP improves oxygenation of the ischemic inner retinal layers by destroying metabolically highly active photoreceptor cells resulting in higher retinal oxygen tension which leads to retinal arteries constriction by autoregulation and in reduction in blood flow(9) Studies suggests that gradual recovery is linked with the restoration of the outer retinal layer after photocoagulation (10-13)
Limited prospective literature exists about scar expansion rate.(14-16) Higaki et al. compared this subject between conventional laser and short pulse pattern scan laser using fundus autofluorescence and a digital image analysis software, finding greater expansion rate in the conventional laser at 12 months of follow up (1.39 ± 0.11 vs. 1.13 ± 0.05, p <0.01). As far we know, there is no comparative data of the scar expansion rate using OCT between both laser groups. Our data shows not only greater initial scar size but greater mean horizontal expansion at 12th week in the subjects who received single spot argon laser PRP (72.12μm ± 61.31μm vs. 43.83μm ± 31.60μm, p<0.001) (Figure 7). Collateral damage to photoreceptors and subsequent apoptosis of surrounding horizontal and amacrine cells occurs after conventional laser although the main target is the RPE. As a consequence, the scar expands, not so when short-pulse laser is performed, in which the retinal damage is limited. (14) Interestingly, we report an initial scar contraction in the conventional laser group. Studies in rabbits demonstrate that in intense and moderate lesions, gliotic scar filling stabilizes after one month, and the wound contracts to approximately 40% of the original lesion diameter with a decrease in early retinal edema(17)
There were several limitations that need to be acknowledged in the current study. First, we had strict inclusion and exclusion criteria, so caution is advised to extrapolate this data to the general population. Second, we had a small sample size. To evaluate the exact changes in CRT, ONH flow and scar expansion rate comparing conventional single spot argon laser with short pulse pattern scan laser, a larger sample size would be warranted, although this demonstrated data may be the beginning of future research.
Conclusión
Non statistically significant changes were noted in BCVA, chromatic vision or contrast sensitivity at the end of the follow up. Short pulse pattern scan laser is more tolerable to patients. There was a significant augmentation of CRT in the single spot argon laser group. The ONH vascular density was restored at 12th week after PRP. There was a greater initial size and horizontal scar expansion in the subjects that received conventional argon laser.